Leading Supplier of Optical & Educational from China About Us Contact Us
ID    PW      New Register    Forget Password
        
 
A13.1018
Metallurgical Microscope, Transmit & Reflect Light
 
A13.1017
Metallurgical Microscope, Reflect Light
 
A13.2603
Metallurgical Microscope
 
A13.2605
Metallurgical Microscope
 
A13.2604-B
Metallurgical Microscope, Reflect & Transmit
 
A13.2604-A
Metallurgical Microscope, Reflect
Home >> Optical Instrument >> Biological Microscope >> Metallurgical Microscope

A13.2502 Portable Measuring Microscope
click to see big img
 
  Features:

A13.2502 Portable Measuring Microscope, is designed to quickly measuring the diameter of power particle, hardness test spot diameter or thickness of coating layer. It is with compact and simple design but very useful for outdoor work with LED light and small size. It can be used for biology medicine agriculture and industry area widely. It is the ideal instrument in medical treatment, teaching demonstration and scientific research.


--WF20x or WF10x Wide Field Micrometer Eyepiece
--Fixed Micrometer With Main Scale In Eyepiece : 0~8mm
--Moving Micrometer With Corss & Double Lines, Moved By Measuring Knob
--Measuring Range 4mm, Division 0.01mm/0.005mm

 
  Detail Specification:
A13.2502 A13.2502-B
Total Magnification 20x 20x
Wide Field Eyepiece

WF20x

WF10x
Objective 1x 2x
Measuring Knob Division 0.01mm Division 0.005mm
Eyepiece Micrometer --Fixed Micrometer With Main Scale In Eyepiece : 0~8mm
--Moving Micrometer With Corss & Double Lines, Moved By Measuring Knob
Measuring Range 4mm
Measuring Precision +/- 0.01mm
Function Measuring Diameter Of Power Particle
Measuring Hardness Test Spot Diameter
Measuring Thickness Of Coating Layer
Measuring All Kinds Of Lenght On Surface...
Theory

As Structure Picture shown, there are 2 micrometers in the microscope:
--Fixed Micrometer With Main Scale In Eyepiece : 0~8mm
--Moving Micrometer With Corss & Double Lines, Moved By Measuring Knob

The Fixed Micrometer has a main scale 0~8mm length, which is used to fix the measurement position. When you turn the Measuring Knob, it will move the Moving Micrometer and show its moving distance on the knob by 0.01mm division. Keep record of its moving distance, it will be the length you need to measure.

Operation Manual

Measuring Diameter Of Power Particle (See Picture)
1. Put the microscope on the surface of objects, turn on the LED light.
2, Adjust eyepiece (2) to see both micrometer clearly in view field.
3. Adjust focusing ring (7) to see objects clearly in view field.
4. Fix the focusing by screw (9).
5. Turn Measuring Knob to move the micrometer, until the cross center locate on left side of the particle. Read out the main scale as 5 and measuring knob scale as 0.36 (a). (Left Picture)
6. Turn Measuring Knob to move the micrometer, locate the cross center on right side of the particle. Read out the main scale as 4 and measuring knob scale as 0.51 (b). (Right Picture)
7. Moving distance (Y) is the difference between (a) and (b) divided by objective magnification (X) : Y=(b-a)/X.
8. Measured Diameter = Moving Distance = (5.36mm-4.51mm)/1X=0.85mm

 

 


 

 